Radchex Techincal Specs - X-ray energy dependence: Simulates relative light output of Photostimulatable Phosphor Plate (PSP) within +/- 3% over kVp range of 60 kVp to 120 kVp and a patient equivalent thickness range of 5 cm to 35 cm (within specified operating rates)
- Digital range: Computed Radiography Light Units; CRLU 0 to 5000.0
- Minimum CRLU Rate: 7/sec (approx. 0.7 mR/sec entrance exposure rate)
- Maximum CRLU Rate: 25000/sec (approx. 2500 mR/sec entrance exposure rate)
- Power On/Off: Manual switch
- Power requirements: Built-in NiMH rechargeable battery pack (9.6 V)
- Typical battery life between charging: Greater then 20 hrs
- Operating environment: 15 degrees C to 35 degrees C (59 to 95 degrees F)
- X-ray beam filter: 1.5 mm Copper (B152-110); 6” by 6” with Velcro straps to attach to x-ray tube collimator housing
- Electronic Cassette: Dimensions 10” be 12” by 0.5” (24 cm x 30 cm x 1.3 cm)
- Weight 1.8 kg (3.9 lb)
Mammochex Meter Technical Specs- X-ray energy dependence: Simulates relative light output of Photostimulatable Phosphor Plate (PSP) within +/- 3% over kVp range of 20kVp to 32kVp and a patient equivalent thickness range of 2cm to 8cm (within specified operating rates)
- Digital Range: Mammo Light Units(MamLU); MamLU: 0 to 6553.5
- Minimum MamLU Rate: 15.0/Sec (approx. 1.5 mR/Sec cassette entrance exposure rate)
- Maximum MamLU Rate: 3000/Sec (approx. 300 mR/Sec cassette entrance exposure rate)
- Power On/Off: Manual Switch
- Power requirements: Built in NiMH rechargeable battery pack (9.6V)
- Operating environment: 15°C to 35°C (59°F to 95°F)
- X-ray beam filter: 2mm Al (type 1100) supplied for plate reader calibration assessment – 4”x 4” to attach to collimator housing
- Electronic Cassette Dimensions: Fits 18 x 24 cm Bucky; Fits 24 x 30 cm Bucky with L adaptor
- Electronic Cassette Weight: 1.8kg (3.9 lb);L adaptor 0.73kg (1.6lb)
|
|
|